Monochromator
Cat. No. Model Spectral Bandwidth Cuvette
S23-3269-030 OPTIZEN NanoQ 1.0 nm (FWHM at Hg 253.7 nm) X
S23-3269-035 OPTIZEN NanoQ Plus O
Cat. No. S23-3269-030 S23-3269-035
Model OPTIZEN NanoQ OPTIZEN NanoQ Plus
Photometrics System Microvolume Spectrophotometer
Light Source Xenon flash lamp
Spectral Bandwidth 1.0 nm (FWHM at Hg 253.7 nm)
Wavelength Range 190 to 850 nm
Wavelength Display(setting) 1 nm
Wavelength Accuracy ¡¾ 1 nm
Life Time Up to 10 years
Detector CMOS linear image sensor (2048 pixels)
Microvolume Minimum Sample Volume 1 ¥ìL
Photometric Range 0.02 - 330 A (10 mm equivalent)
Detection Limit 2 ng/¥ìL (dsDNA)
0.06 mg/mL (BSA)
0.003 mg/mL (IgG)
Maximum Concentration 16,500 ng/¥ìL (dsDNA), 400 mg/mL (BSA)
Cuvette Photometric Range - 0 - 2 A
Detection Limit - 0.2 ng/¥ìL (dsDNA)
0.006 mg/mL (BSA)
0.0003 mg/mL (IgG)
Center Height (Z-height) - 15 mm
Heating (Optional) - 37 ¡ÆC
Absorbance Precision 0.002 A (0.5 mm path) or 1%
Absorbance Accuracy 3% (at 0.97A at 302 nm)
Measurement Time < 8 seconds
Display 7 inch color LCD with touch screen
Operating System(OS) Android¢â
Software Compatibility Windows 7 and 10
Weight(kg) 3 kg
Footprint(W£øD) 216 x 290 mm